- Hoge kwaliteit en uitgebreide service
- Al meer dan 100 jaar een begrip
- Een breed assortiment
Orbitale schudbeweging amplitude 2mm
Schudsnelheid 250–1400 rpm
Afmetingen (bxdxh) 205x230x130 mm
Gewicht 4 kg
Levering zonder verwarmingsblokken
Merk | BioSan |
---|---|
Afscheider zuigzijde | Nee |
Condenser uitlaat | Nee |
BioSan TS-100 Thermo Schudder voor Micro testtubes
Thermo–Shaker TS-100 provides intensive mixing and temperature control of samples in microtest tubes or PCR plate. Functions of heating and mixing can be performed both simultaneously and independently, i.e. the unit implements three devices in one:
1. Shaker;
2. Dry–block Thermostat;
3. Thermo–Shaker.
TS-100 is used for DNA analysis sample preparation, for extraction of proteins, polysaccharides, lipids and other cellular components.
Heating source is a printed heating board (12 V). Mixing is provided by movement of orbital type.
The instrument is applicable in:
- DNA analysis — DNA extraction;
- Biochemical study of enzymatic reactions and processes;
- Extraction of metabolites from cellular material.
Temperature Calibration Function
With the help of the temperature calibration function the user can calibrate the unit approx. ±6% of the selected temperature to compensate differences in the thermal behaviour of tubes from different manufacturers.
Specification
Temperature setting range | +25°C ... +100°C |
Temperature control range | 5°C above ambient ... +100°C |
Temperature setting resolution | 0.1°C |
Temperature stability | ±0.1°C |
Temperature accuracy at +37°C | ±0.5°C |
Average heating speed from +25°C to +100°C | 4°C/min |
Temperature uniformity over the block | -> |
at +37°C | ±0.1°C |
at +60°C | ±0.2°C |
at +100°C | ±0.2°C |
Temperature calibration coefficient range | 0.936...1.063 (± 0.063) |
Speed control range | 250–1400 rpm |
Acceleration time | 3 sec |
Orbit | 2 mm |
Microprocessor controlled temperature, mixing speed and operation time | + |
Display | LCD, 16 x 2 signs |
Digital time setting | 1 min–96 hrs (1 min increment) |
Maximum continuous operation time | max. 96 hours |
Overall dimensions | 205x230x130 mm |
Weight | 3.2 kg |
Input current/power consumption | 12 V, 3.5 A / 42 W |
External power supply | Input AC 100–240 V; 50/60 Hz; Output DC 12 V |